9020 Multifunction Calibrator
9 ppm AC/DC multifunction calibrator for multimeters , power meters and generally all kinds of electric meters. Implementing modular design, 9010+ can be further customized to calibrate oscilloscopes, insulation testers, sources and transducers. Multifunction calibrator 9020 is designed as universal calibration tool for electrical calibration laboratories, covering most of their workload like 6½ digit multimeters, clamp meters, ohm meters, power meters and power analysers, energy meters, transducers, insulation testers, process meters, oscilloscopes and many others. High load capacity of voltage output (up to 50 mA) allows for calibration of high-consumption analogue meters. Installed harmonic and non-harmonic shape signals allow for testing meter sensitivity to distorted signals by a signal with various crest factor. 9020 takes advantage of 9010's wide workload coverage and modern user interface ans pushes its performance even further by accuracy improvements in most functions, current range extension to 30 A, power range extension to 1575 kW and ultimately 1.1 GHz o-scope calibration capability with SC1 option.
Key Specifications
| Parameter | Specification |
|---|---|
| DC Voltage |
0 - 1050 V ± 9 ppm / year |
| AC Voltage |
1 mV - 1050 V, 3 Hz - 300 kHz Sinusoidal & Non-Sinusoidal waveforms, harmonics ± 160 ppm / year |
| DC Current |
0 - 30 A (1500 A with 0950 Current Coil) ± 100 ppm / year |
| AC Current |
10 µA - 30 A, 15 Hz - 30 kHz (1500 A with 0950 Current Coil)
Sinusoidal & Non-Sinusoidal waveforms, harmonics ± 300 ppm / year |
| DC Power & Energy |
40 µW - 31.5 kW (1575 kW with 0950 Current Coil) ± 160 ppm / year |
| AC Power & Energy |
40 µW - 31.5 kW, 15 - 1000 Hz (1575 kW with 0950 Current Coil)
50 harmonic products, voltage from current ± 670 ppm / year |
| Resistance |
0 Ω - 1 GΩ, resolution 6½ digits ± 15 ppm / year |
| Insulation Resistance |
100 kΩ - 100 GΩ Rated up to 1.5 kV |
| Capacitance |
600 pF - 120 mF, resolution 6½ digits ± 0.15 % / year |
| Temperature |
TC BCDEG2JKMNRST with cold junction compensation RTD ± 0.02°C / year |
| Frequency |
0.1 Hz - 2 MHz ± 5 ppm / year |
Technical Specifications
| Parameter | Specification |
|---|---|
| DC Voltage |
0 – 12 V mVDC resolution down to 10 nV (±50 ppm/year) |
| DC Current |
0 – 24 mA (±150 ppm/year) |
| Resistance |
0 Ω – 20 kΩ Resolution 6½ digits (±200 ppm/year) |
| Temperature |
TC BCDEG2JKMNRST RTD (±0.08°C/year) |
| Frequency |
1 Hz – 100 kHz (±50 ppm/year) |