9020 Multifunction Calibrator

9 ppm AC/DC multifunction calibrator for multimeters , power meters and generally all kinds of electric meters. Implementing modular design, 9010+ can be further customized to calibrate oscilloscopes, insulation testers, sources and transducers. Multifunction calibrator 9020 is designed as universal calibration tool for electrical calibration laboratories, covering most of their workload like 6½ digit multimeters, clamp meters, ohm meters, power meters and power analysers, energy meters, transducers, insulation testers, process meters, oscilloscopes and many others. High load capacity of voltage output (up to 50 mA) allows for calibration of high-consumption analogue meters. Installed harmonic and non-harmonic shape signals allow for testing meter sensitivity to distorted signals by a signal with various crest factor. 9020 takes advantage of 9010's wide workload coverage and modern user interface ans pushes its performance even further by accuracy improvements in most functions, current range extension to 30 A, power range extension to 1575 kW and ultimately 1.1 GHz o-scope calibration capability with SC1 option.

Key Specifications

Parameter Specification
DC Voltage 0 - 1050 V
± 9 ppm / year
AC Voltage 1 mV - 1050 V, 3 Hz - 300 kHz
Sinusoidal & Non-Sinusoidal waveforms, harmonics
± 160 ppm / year
DC Current 0 - 30 A (1500 A with 0950 Current Coil)
± 100 ppm / year
AC Current 10 µA - 30 A, 15 Hz - 30 kHz (1500 A with 0950 Current Coil)
Sinusoidal & Non-Sinusoidal waveforms, harmonics
± 300 ppm / year
DC Power & Energy 40 µW - 31.5 kW (1575 kW with 0950 Current Coil)
± 160 ppm / year
AC Power & Energy 40 µW - 31.5 kW, 15 - 1000 Hz (1575 kW with 0950 Current Coil)
50 harmonic products, voltage from current
± 670 ppm / year
Resistance 0 Ω - 1 GΩ, resolution 6½ digits
± 15 ppm / year
Insulation Resistance 100 kΩ - 100 GΩ
Rated up to 1.5 kV
Capacitance 600 pF - 120 mF, resolution 6½ digits
± 0.15 % / year
Temperature TC BCDEG2JKMNRST with cold junction compensation
RTD ± 0.02°C / year
Frequency 0.1 Hz - 2 MHz
± 5 ppm / year

Technical Specifications

Parameter Specification
DC Voltage 0 – 12 V
mVDC resolution down to 10 nV (±50 ppm/year)
DC Current 0 – 24 mA
(±150 ppm/year)
Resistance 0 Ω – 20 kΩ
Resolution 6½ digits (±200 ppm/year)
Temperature TC BCDEG2JKMNRST
RTD (±0.08°C/year)
Frequency 1 Hz – 100 kHz
(±50 ppm/year)